User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: PM 5135 (Calibration) Ver 1.0 DATE: 24-Oct-93 AUTHOR: User Contributed REVISION: 1.0 ADJUSTMENT THRESHOLD: 75% NUMBER OF TESTS: 30 NUMBER OF LINES: 278 CONFIGURATION: Fluke 8842A CONFIGURATION: Philips PM 5191 CONFIGURATION: Philips PM 3365A CONFIGURATION: Philips PM 6654C CONFIGURATION: HP 8568A ============================================================================= * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Source: Service Manual PM 5135 P/N: 9499 455 00511 @ 910721 * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * TUR Configuration: * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * When the T.U.R. is < 2:1, the source instrument, e.g. 5700A, must be * characterized and appropriate corrections entered in the correction file * (e.g. 5700.COR). Where T.U.R.'s are > 2:1 but < 4:1, the test tolerance * is reduced by root sum square (rss) techniques. * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * This procedure is not executable in MET/CAL versions below ver. 2.00 * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Fluke make no warranty, expressed or implied, as to the fitness or * useability of this procedure. * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * Main Procedure TESTS: Calibration * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ * The following subprocedures are called from this main procedure: * 1 SUB PART1 PM 5135 (CALIBRATION) * TESTS/LINES : 32 / 249 #REV :1.0 * 2 SUB PART2 PM 5135 (CALIBRATION) * TESTS/LINES : 9 / 76 #REV :1.0 * ÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄÄ STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P F W 1.002 HEAD INTRODUCTION 1.003 DISP This procedure is intended to: 1.003 DISP - Perform re-calibration to manufactures specification. 1.003 DISP 1.003 DISP NOTE: 30 Minutes warmup is required before testing 1.004 HEAD {7.2.1 Supply Voltages} 1.005 DISP Connect the 8840 to C180 on the side closest to IC123. 1.005 DISP Connect the 8840 ground to the UUT chassis. 1.006 RSLT =Step 1 1.007 8842 15.00V 0.6U 2W 2.001 DISP Connect the 8840 to C181 on the side closest to IC123. 2.001 DISP Connect the 8840 ground to the UUT chassis. 2.002 RSLT =Step 2 2.003 8842 -15.00V 0.6U 2W 3.001 DISP Connect the 8840 to C182 on the side closest to IC124. 3.001 DISP Connect the 8840 ground to the UUT chassis. 3.002 RSLT =Step 3 3.003 8842 22.00V 0.9U 2W 4.001 DISP Connect the 8840 to C183 on the side closest to IC124. 4.001 DISP Connect the 8840 ground to the UUT chassis. 4.002 RSLT =Step 4 4.003 8842 -22.00V 0.9U 2W 5.001 HEAD {7.2.2 Sweep }Referance Voltage 5.002 DISP Connect the 8840 to TP104. (IC101 Pin 14) 5.002 DISP Connect the 8840 ground to the UUT chassis. 5.003 RSLT =Step 5 5.004 8842 10.000V 0.002U 2W 5.005 MESS Adjust VR101 for 10.000 +/- 0.002 VDC. 5.006 8842 V S 2W 5.007 MESS 5.008 JMP 5.004 6.001 HEAD 7.2.2 Sweep Time 6.002 IEEE [@8842]*[GTL] 6.003 DISP Setup UUT as follows: 6.003 DISP - SWEEP TIME: Right Stop 6.003 DISP 6.003 DISP Connect the 8840 (+ lead) to TP101 on the side closest 6.003 DISP to IC101. 6.003 DISP Connect the 8840 (- lead) to TP101 on the side closest 6.003 DISP to C105. 6.004 RSLT =Step 6 6.005 8842 -69.000mV 2U 2W 6.006 MESS Adjust VR508 (Unit 2) for -69.0 +/- 2.0 mVDC 6.007 8842 V S 2W 6.008 MESS 6.009 JMP 6.005 7.001 HEAD 7.2.2 Sweep Control Elements 7.002 IEEE [@3365][SPL]SET STANDARD,VER A,ATT 2E+00,CPL DC[10] 7.002 IEEE POS -1000,MSC AUX,MEM ON[10] 7.002 IEEE HOR MTB,FCN ON,TIM 20E-09,TRG AUT,TSO A,TSL POS[10] 7.002 IEEE SPL CURSOR,FCN ON,FIRST 1000,SECOND 1001,RMS ON[10] 7.003 DISP Setup UUT as follows: 7.003 DISP - All three sweep buttons not pressed. 7.003 DISP 7.003 DISP Connect the 3365 to the UUT SWEEP OUTPUT (Rear Panel). 7.003 DISP Using a bnc cable. 7.004 IEEE [@3365][SPL]VER A,CPL ZERO,SPL CURSOR,RMS ?[10][I] 7.004 IEEE VER A,CPL DC,SPL CURSOR,RMS ?[10][I] 7.005 RSLT =Step 7 7.006 MEMC 00.400V +2U 8.001 DISP Press the SWEEP ON button 8.002 IEEE [@3365][SPL]SPL CURSOR,RMS ?[10][I] 8.003 RSLT =Step 8 8.004 MEMC 09.400V -6U 9.001 DISP Press the TRIG button. 9.002 IEEE [@3365][SPL]SPL CURSOR,RMS ?[10][I] 9.003 RSLT =Step 9 9.004 MEMC 00.400V +2U 10.001 RSLT =LED turn ON? 10.002 EVAL Did the LED turn ON? 11.001 DISP Set the SWEEP TIME to the first mark (3 Seconds) 11.001 DISP and press the TRIG button. 11.002 IEEE [@3365][D5000][SPL]SPL CURSOR,RMS ?[10][I] 11.003 RSLT =Step 10 11.004 MEMC 09.400V -6U 12.001 RSLT =LED turn OFF? 12.002 EVAL Did the output Rise to 10VDC and the LED turn off? 13.001 MESS Press the CONT button. 13.002 RSLT =Cont Sweep from 0 to 10V then reset,LED off? 13.003 EVAL Does the output Cont Sweep from 0 to 10V then reset,LED off? 14.001 MESS 14.002 DISP Release the CONT button. 14.003 IEEE [@3365][D2000][SPL]SPL CURSOR,RMS ?[10][I] 14.004 RSLT =Step 12 14.005 MEMC 09.400V -6U 15.001 DISP Release the SWEEP ON button. 15.002 IEEE [@3365][SPL]SPL CURSOR,RMS ?[10][I] 15.003 RSLT =Step 13 15.004 MEMC 00.400V +6U 16.001 HEAD {7.2.3 Frequency} Comparator Supply 16.002 DISP Setup UUT as follows: 16.002 DISP - FREQ RANGE: x100, FREQ DIAL: 100 (10 KHz) 16.002 DISP - FREQ OFFSET: Centered (Detent), WAVEFORM: Triangle 16.002 DISP 16.002 DISP Disconnect cable to PL106 16.002 DISP 16.002 DISP Connect the 8840 to Unit 1, PL106, Pin 5. 16.002 DISP Connect the 8840 ground to the UUT chassis. 16.003 RSLT =step 14 16.004 8842 10 0.000mV 5.000U 2W 16.005 MESS Adjust VR109 for 0 +/-5 mVDC. 16.006 8842 V S 2W 16.007 MESS 16.008 JMP 16.004 17.001 HEAD 7.2.3 Frequency Ranges - Commutator Balance 17.002 IEEE [@3365][SPL]SET STANDARD,VER A,ATT .1E+00,CPL AC, 17.002 IEEE MSC AUX,MEM ON, 17.002 IEEE SPL CURSOR,FCN ON,FIRST 1000,SECOND 3000,PEAK ON, 17.002 IEEE HOR MTB,FCN ON,TIM .1E-03,TRG AUT,TSO A,TSL POS[10][GTL] 17.003 DISP Setup UUT as follows: 17.003 DISP - FREQ RANGE: x10k, FREQ DIAL: 0.2 (2 KHz) 17.003 DISP - FREQ OFFSET: Centered (Detent) 17.003 DISP 17.003 DISP Connect the 3365A to TP102. Use a 10:1 probe. 17.003 DISP 17.003 DISP Adjust VR106 for Minimum square wave. 17.003 DISP NOTE: Adjust Ch A Sensativity as required. 17.004 IEEE [@3365][SPL]SPL CURSOR,PEAK ?[10][I][GTL] 17.005 RSLT =Step 15 17.006 MEMC 00.400V +6U 18.001 DISP Connect the 3365A to TP103. Use a 10:1 probe. 18.001 DISP 18.001 DISP Adjust VR107 for Minimum square wave. 18.001 DISP NOTE: Adjust Ch A Sensativity as required. 18.002 IEEE [@3365][SPL]SPL CURSOR,PEAK ?[10][I][GTL] 18.003 RSLT =Step 16 18.004 MEMC 00.400V +6U 19.001 HEAD 7.2.3 Frequency Ranges 19.002 DISP Setup UUT as follows: 19.002 DISP - FREQ RANGE: x100, FREQ DIAL: Right Stop (<0.1) 19.002 DISP - FREQ OFFSET: Centered (Detent), WAVEFORM: Triangle 19.002 DISP - SWEEP: ON (LED off, if not press TRIG) 19.002 DISP 19.002 DISP Connect the 8840 to the UUT SWEEP OUTPUT (Rear Panel) 19.003 RSLT =Step 17 19.004 8842 10.000V 0.013U 2W 20.001 DISP Setup UUT as follows: 20.001 DISP - SWEEP STOP: Left Stop (>200) 20.001 DISP 20.001 DISP Connect the 6654 Ch A (Counter) to PL106, Pin 5. 20.001 DISP Use a 10:1 probe. 20.002 IEEE [@6654]D MS1 F1[I][D100]X[I][D100]X[I] 20.003 MEM/ 1000 20.004 RSLT =Step 18 20.005 MEMC 20.000KH 0.2U 20.006 DISP Setup UUT as follows: 20.006 DISP - SWEEP STOP: Right Stop (200) 20.006 DISP 20.006 DISP Adjust VR108 and/or VR104 for 20 +/-.2 KHz. 20.006 DISP - SWEEP STOP: Left Stop (0.1) 20.006 DISP 20.006 DISP Adjust VR108 and/or VR104 for 10 +/-.1 Hz. 20.006 DISP Repeat as nessecary balance adjustments. 20.007 JMP 20.001 21.001 DISP Setup UUT as follows: 21.001 DISP - SWEEP STOP: Left Stop (0.1) 21.002 IEEE [@6654]D MS1 F1[I][D100]X[I][D100]X[I] 21.003 RSLT =Step 19 21.004 MEMC 10.000H 0.1U 21.005 JMP 20.006 22.001 DISP Setup UUT as follows: 22.001 DISP - SWEEP: OFF 22.002 RSLT =Step 20 22.003 8842 10 0.000mV 13.000U 2W 23.001 IEEE [@8842]*[GTL] 23.002 DISP Setup UUT as follows: 23.002 DISP - FREQUENCY: 200 (20 KHz), SWEEP: Off 23.002 DISP 23.002 DISP Connect the 8840 to Unit 2, PL501, Pins 1 and 4 23.002 DISP (Potmeter Board) 23.002 DISP 23.002 DISP NOTE: Do Not disconnect PL501 from Unit2. 23.003 IEEE [@6654]D MS1 F1[I][D100]X[I][D100]X[I] 23.004 MEM/ 1000 23.005 RSLT =Step 21 23.006 MEMC 20.000KH 0.2U 23.007 DISP Adj VR501 (Unit 2) for 20.00 +/- .60KHz. 23.008 DISP Setup the UUT as Follows: 23.008 DISP SWEEP: On 23.008 DISP Adj VR502 (Unit 2) for 0.00 +/- 10mVDC. 23.009 DISP Setup tht UUT as Follows: 23.009 DISP SWEEP: Off 23.010 JMP 23.003 24.001 DISP Setup UUT as follows: 24.001 DISP - SWEEP: On 24.002 RSLT =Step 22 24.003 8842 10 0.000mV 10.000U 2W 24.004 DISP Setup tht UUT as Follows: 24.004 DISP SWEEP: Off 24.005 JMP 23.007 25.001 DISP Setup UUT as follows: 25.001 DISP - FREQUENCY: Right Stop (<0.1), FREQ RANGE: x1 25.001 DISP - SWEEP STOP: Right Stop (>200), SWEEP: On 25.001 DISP 25.001 DISP Disconnect the 8840 from PL501 25.002 IEEE [@6654]D MS1 F1[I][D100]X[I][D100]X[I] 25.003 RSLT =Step 23 25.004 MEMC 200.00H 2U 25.005 DISP Adj VR103 for 200.00 +/- 2Hz 25.006 JMP 24.003 26.001 DISP Setup UUT as follows: 26.001 DISP - SWEEP STOP: Left Stop (<0.1) 26.001 DISP 26.001 DISP NOTE: Test can take up to 45 seconds. 26.002 IEEE [@6654][T0]D AC0 MS1 F3[I]X[I] 26.003 RSLT =Step 24 26.004 MEMC 10.000S -100% 26.005 DISP Setup UUT as follows: 26.005 DISP - SWEEP STOP: Right Stop (>200) 26.006 JMP 25.002 27.001 DISP Setup UUT as follows: 27.001 DISP - FREQ RANGE: x0.01, SWEEP STOP: Right Stop (>200) 27.002 IEEE [@6654][T0]D MS1 F3[I]X[I] 27.003 RSLT =Step 25 27.004 MEMC 00.500S 0.005U 27.005 DISP Adj VR102 for 0.5000 +/- 0.005S 27.006 JMP 27.002 28.001 DISP Setup UUT as follows: 28.001 DISP - FREQ RANGE: x10k, SWEEP STOP: Right Stop (>200) 28.002 IEEE [@6654]D MS1 F1[I][D100]X[I][D100]X[I] 28.003 MEM/ 1000000 28.004 RSLT =Step 27 28.005 MEMC 02.000MH 0.06U 28.006 DISP Adj VR105 for 2.000 +/- .06MHz. 28.007 JMP 28.002 29.001 DISP Setup UUT as follows: 29.001 DISP - SWEEP STOP: Left Stop (<0.1) 29.002 IEEE [@6654]D MS1 F1[I][D100]X[I][D100]X[I] 29.003 MEM/ 1000 29.004 RSLT =Step 28 29.005 MEMC 01.000KH +100% 29.006 DISP Setup tht UUT as Follows: 29.006 DISP SWEEP: Off 29.007 JMP 28.006 30.001 CALL Sub Part1 PM 5135 (Calibration) 30.002 IEEE [@8568][GTL] 30.003 5191 * S 30.004 END